0

Description

This is a flexible protocol accommodating a wide range of commercial RNA extraction workflows for COVID-19 sample processing. The protocol is broken down into 5 main parts:

  • binding buffer addition to samples
  • bead wash 3x using magnetic module
  • final elution to chilled PCR plate

Lysed samples should be loaded on the magnetic module in a NEST 96-deepwell plate. For reagent layout in the 2 12-channel reservoirs used in this protocol, please see "Setup" below.

For sample traceability and consistency, samples are mapped directly from the magnetic extraction plate (magnetic module, slot 4) to the elution PCR plate (temperature module, slot 1). Magnetic extraction plate well A1 is transferred to elution PCR plate A1, extraction plate well B1 to elution plate B1, …, D2 to D2, etc.

Explanation of complex parameters below:

  • park tips: If set to yes (recommended), the protocol will conserve tips between reagent addition and removal. Tips will be stored in the wells of an empty rack corresponding to the well of the sample that they access (tip parked in A1 of the empty rack will only be used for sample A1, tip parked in B1 only used for sample B1, etc.). If set to no, tips will always be used only once, and the user will be prompted to manually refill tipracks mid-protocol for high throughput runs.
  • track tips across protocol runs: If set to yes, tip racks will be assumed to be in the same state that they were in the previous run. For example, if one completed protocol run accessed tips through column 5 of the 3rd tiprack, the next run will access tips starting at column 6 of the 3rd tiprack. If set to no, tips will be picked up from column 1 of the 1st tiprack.
  • flash: If set to yes, the robot rail lights will flash during any automatic pauses in the protocol. If set to no, the lights will not flash.

OT-2

Download Protocol:


Process:

  1. Select your protocol parameters.
  2. Download your protocol.
  3. Upload your protocol into the OT App.
  4. Set up your deck according to the deck map.
  5. Calibrate your labware, tiprack and pipette using the OT App. For calibration tips, check out our support article.
  6. Hit "Run".

Additional Notes:

If you have any questions about this protocol, please contact the Protocol Development Team by filling out the Troubleshooting Survey.